Skip to content Skip to footer

COOK Vacuum Pump

DESCRIPTION:

Designed for aspiration of follicle contents. Equipped with a manual pressure regulator.

Used to provide a low flow, regulated vacuum up to -500 mm Hg for general suction.

  • A pedal activates the vacuum response at the needle tip. The pump maintains a constant vacuum at its set level.
  • The boost button on the front panel increases the vacuum pressure to clear blockages in the ovum aspiration needle.
  • A volume-adjustable tone indicates when vacuum is applied.
  • The LED can display the vacuum pressure in mm Hg and kPa.
  • This pump is not intended for use in emergency airway clearing.

 

ADDITIONAL INFORMATION:

Calibration of the aspiration device: setting the optimal pressure level minimizes damage to oocytes during aspiration of the follicle contents. The optimum for preserving the oocyte-cumulus complex is a volumetric aspiration rate of 20-25 ml/min. To set the pump pressure in mmHg corresponding to this aspiration rate, it is necessary to calibrate the device with the attached line and needle by aspirating water.

Technical characteristics:
Power supply 100 - 240 VAC
The pressure generated From -10 to -500 mmHg
Dimensions WxHxD 200 x 100 x 350 mm
Weight 3.2 kg
Catalog number Description
K-MAR-5200 Aspiration pump, connecting tube with hydrophobic filter, foot pedal for foot control of the pump